Overview
This study aims to test the accuracy of the Perin Health Patch (PHP), a non-invasive, chest-worn device that measures blood oxygen levels, by comparing its readings to a standard pulse oximeter device. The PHP device uses light sensors to measure blood oxygen saturation (SpO2), similar to how standard pulse oximeters work, and could allow for continuous remote monitoring at home. Participants in this study will breathe low-oxygen air through a mask to simulate different oxygen levels in a controlled environment. The goal is to see if the PHP device can accurately measure blood oxygen levels under different conditions and meet safety standards set by the FDA and international guidelines.
Description
The Perin Health Patch (PHP) is a non-invasive, chest-worn medical device that continuously measures blood oxygen saturation (SpO2), alongside other vital signs. The PHP's performance will be assessed under controlled hypoxemia (low oxygen levels), where its readings will be compared to a standard pulse oximeter validated against co-oximeter reference values.
In this single-site, prospective study, up to 25 healthy adult volunteers will undergo controlled desaturation sessions. During each session, a mask delivering a low-oxygen gas mixture will be used to safely lower participants' SpO2 levels. This method simulates a range of oxygenation states, from 70% to 100% SpO2, which is critical for testing the device's accuracy across a clinically relevant range. The PHP's readings will be captured alongside those from two standard pulse oximeters placed on both hands. Participants will be carefully monitored by clinical staff throughout the study, with emergency oxygen and other medical equipment readily available to ensure participant safety.
Participants will complete two desaturation "runs," each lasting approximately 10-15 minutes, with brief breaks between sessions. During some sessions, participants may be asked to walk slowly on a treadmill to simulate movement conditions. This study design aligns with the ISO 80601-2-61 standard, supporting robust accuracy assessments under both stationary and motion conditions.
Eligibility
Inclusion Criteria:
- Adults aged 18 years to 50 years
- The subject is in good general health with no evidence of any medical problems,
- Fully vaccinated for COVID-19
- Fluent in both written and spoken English
- Willing and able to provide informed consent,
- Able to comply with study procedure.
Exclusion Criteria:
- Children (under the age of 18)
- Adults above 50 years old
- History of heart, lung, kidney, or liver disease.
- Obesity (BMI > 30),
- Diagnosis of asthma, sleep apnea, or use of CPAP.
- Diagnosis of Raynaud's disease
- Unacceptable collateral circulation based on an exam by the investigator (Allen's test).
- Pregnant, lactating or trying to get pregnant
- Current smoker
- History of diabetes
- Clotting disorder
- Hemoglobinopathy or history of anemia
- History of fainting or vasovagal response
- Any other serious systemic illness
- Any other condition which, in the opinion of the investigators, would make them unsuitable for the study,
- Any injury, deformity, or abnormality at the sensor sites that in the opinion of the investigators' would interfere with the sensors working correctly,
- Any evidence, in the opinion of the investigators, of medical problems or poor general health
- Any symptoms related to COVID-19
- No vaccinated for COVID-19
- Recent injection with methylene blue
- History of reactions to medical adhesives,
- Unable or unwilling to provide informed consent,
- Inability to comply with the study procedure,
- Non-English Speaker.